• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA FEMUR CEMENTED (PS) STANDARD RIGHT SIZE 6; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA FEMUR CEMENTED (PS) STANDARD RIGHT SIZE 6; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42500606002
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Bursitis (4832)
Date of Event 02/19/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via clinical study that a patient underwent an initial total knee arthroplasty.Approximately four (4) years post-implantation, the patient experienced a sudden onset of stiffness, pain, and bursitis with no apparent cause.The patient was treated with a cortisone injection and the symptoms were reported to have resolved fourteen (14) months post-onset.Due diligence is in progress for this event; to date no additional information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: medical product: tibia cemented 5 degree stemmed right size d: catalog #42532006702, lot #63474954; articular surface fixed bearing posterior stabilized (ps) right 10 mm height: catalog #42521400510, lot #62296922.G2: foreign: netherlands.H6: proposed component code: mechanical (g04) - femur.The product will not be returned to zimmer biomet for investigation, as the product currently remains implanted.The investigation is in process.Upon receipt of additional information or completion of the investigation, a follow-up mdr be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; e1; e2; e3; g2; g3; h1; h2; h3; h6; h10; h11.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Radiographs were provided and reviewed by a health care professional and it was determined further review would not enhance the investigation.Reported event was unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that two months post-resolution of previous incident patient began to again experience discomfort with the implanted prosthesis.Steroid medication was prescribed and no additional intervention was required.All available information has been provided.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following section has been corrected: h11.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Radiographs were provided and reviewed by a health care professional and it was determined further review would not enhance the investigation.Root cause was unable to be determined.Reported event was unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA FEMUR CEMENTED (PS) STANDARD RIGHT SIZE 6
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22871995
Report Number3007963827-2025-00325
Device Sequence Number16024909
Product Code JWH
UDI-Device Identifier00889024229204
UDI-Public(01)00889024229204(17)251231(10)63253075
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date12/31/2025
Device Catalogue Number42500606002
Device Lot Number63253075
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2025
Supplement Date Received by Manufacturer11/11/2025
11/18/2025
Initial Report FDA Received Date08/22/2025
Supplement Report FDA Received Date12/08/2025
12/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-