• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL 4.5HLX ADV BR ANC-DYNA STR/BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL 4.5HLX ADV BR ANC-DYNA STR/BL; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 222001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a rotator cuff repair procedure; it was discovered that the anchor on the 4.5hlx adv br anc-dyna str/bl device broke off.All the broken parts were removed.It was not reported if there were any delays to the surgical procedure.It was reported that a spare device was used to complete the surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: e1: the reporter¿s complete facility address was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: update: d4, h4: expiration date, primary udi number and device manufacture date added.Investigation summary: the product was not returned to j&j medtech orthopaedics; however, a photo was provided for evaluation.Visual inspection of the returned photo found that the device is shown in used condition.The anchor was found broken.No other anomalies were noted.The overall complaint was confirmed as the observed condition of the 4.5hlx adv br anc-dyna str/bl would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to procedural variables, such handling of the device or product interaction during procedure; axial misalignment occurred and consequently the anchor fracture.As per ifu 116034; inserting the awl or drill to less than the specified depth, axial misalignment, or levering with the anchor upon insertion may result in anchor fracture.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history record review: a device history review was performed and no non-conformances were identified related to the reported condition.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: correction: d9: corrected.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11: additional narrative: correction: d9: corrected.Investigation summary- the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection found that the device had the anchor still attached to the shaft and broken at the distal end.The anchor had foreign matter, presumably biological matter.No other anomalies were noted.The overall complaint was confirmed as the observed condition of 4.5hlx adv br anc-dyna str/bl would have contributed to the complained device issue.¿ based on the investigation findings, the potential cause is traced to the procedural variables, such handling of the device or product interaction during procedure.The possible root cause can be associated with improper instrument use, axial misalignment or levering during insertion.As per instructions for use (ifu) inserting the awl or drill to less than the specified depth, axial misalignment, or levering with the anchor upon insertion may result in anchor fracture.2) do not apply a bending force to the inserter.This can damage the anchor or inserter tip.Properly handling and attention to the approved use of the device diminishes the risk of failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5HLX ADV BR ANC-DYNA STR/BL
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key22872895
Report Number1221934-2025-03222
Device Sequence Number20255895
Product Code MAI
UDI-Device Identifier10886705029402
UDI-Public(01)10886705029402(10)241L284(17)260930
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K173859
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 08/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Catalogue Number222001
Device Lot Number241L284
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Supplement Date Received by Manufacturer08/25/2025
09/11/2025
09/12/2025
09/16/2025
Initial Report FDA Received Date08/22/2025
Supplement Report FDA Received Date09/02/2025
09/11/2025
09/15/2025
09/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-