| Model Number |
WA51181L |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Poor Visibility (4072)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject device had the suction/irrigator handle getting stuck in the suction position during an abdominal laparoscopic procedure, causing the co2 in the patient's abdomen to get sucked out.This made the surgeons lose visibility multiple times.It also reduced surgical efficiency as they had to keep pausing to allow the abdomen to reinflate.There was minimal delay to the laparoscopic roux-en-y gastric bypass procedure.The device was replaced with a similar device.There were no reports of patient harm or serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updates fields: b5, d9, g3, h2, h3, h6 and h11.The device was returned to olympus for inspection.Based on the results of the investigation the customer allegation was confirmed.The irrigation/suction tube was attached and upon pressing on the irrigation/suction levers it was confirmed that the suction lever was stuck and not retracting back into its normal position.It was likely due to a component problem.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Search Alerts/Recalls
|