ALCON LABORATORIES IRELAND LTD. CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CCA0T0
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that during the intraocular lens (iol) implant procedure, plunger slid past iol prior to lens insertion.The new lens got implanted and the procedure was completed.Additional information has been received stating that faulty lens while loading device with confirmed no contact with patient.Additional information has been requested however no further information available.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6., and h.11.The product was returned for analysis, and the reported complaint was observed.The device was returned loose in the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been advanced to mid nozzle and is advanced under the intraocular lens (iol).The iol is advanced to mid nozzle and is damaged.We are unable to determine the root cause for the reported complaint "faulty lens while loading device".Plunger underride was observed.Plunger underride may occur: if the lens has become misaligned in the lens bay during the manufacturing process.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ophthalmic visco surgical device (ovd) may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to underride the lens.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC UV ABSORBING IOL WITH AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22874449
Report Number9612169-2025-01696
Device Sequence Number10371449
Product Code HQL
UDI-Device Identifier00380652397193
UDI-Public(01)00380652397193(17)271108(11)231110(10)25741722
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 11/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCA0T0
Device Lot Number25741722
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2025
Supplement Date Received by Manufacturer10/21/2025
Initial Report FDA Received Date08/22/2025
Supplement Report FDA Received Date11/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/10/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC VISCOELASTIC SYSTEM
Patient SexUnknown
-
-