| Catalog Number |
230H4001 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference report: 3012447612-2025-00296.
|
| |
|
Event or Problem Description
|
|
It was reported that two tether set screwdriver tips fractured during a surgery.The c-rings broke off in the tulip head of the tether screws.There was no patient impact; however, there was a 30-45 minute delay to retrieve the broken pieces.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tip has fractured.Root cause: this event could possibly be attributed to excessive forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that two tether set screw driver tips fractured during a surgery.The c-rings broke off in the tulip head of the tether screws.There was no patient impact; however, there was a 30-45 minute delay to retrieve the broken pieces.This is report two of two for this event.
|
| |
|
Search Alerts/Recalls
|