• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC ACROBAT-I STABILIZER Z; STABILIZER, HEART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC ACROBAT-I STABILIZER Z; STABILIZER, HEART Back to Search Results
Model Number OM-10000Z
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during a coronary artery bypass procedure, acrobat-i stabilizer z handle was inoperable and wouldn't secure the stabilizer; it kept rotating without tightening.To avoid further delays in the surgery, it was replaced with another one that worked correctly.No injuries or required interventions.
 
Additional Manufacturer Narrative
(b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).Updated sections: a2, a3a, a4, b4, g3, g6, h2, h11.The hospital reported that during a coronary artery bypass procedure, acrobat-i stabilizer z handle kept rotating without tightening.To avoid delays in the surgery, it was replaced with another one that worked correctly.No injuries or required interventions.The investigation has been started since the complaint was received, the following contents have been conducted: 1.Dhr review: the dhr of the reported lot has been reviewed.No non-conformity relevant with the reported issue is observed.All the products had been performed 100% mechanical function test during production.They all passed the function test which demonstrate the knob can be tightened and can also tighten the flexlink arm.2.Trend review: in the last 12 months, the occurrence rate is found to be 0.074% which is under anticipated occurrence rate.3.Device was not received for evaluation, the specific root cause for this failure is impossible to define.4.Review complaint historical data, the failure knob difficult/ unable to tighten-arm does not lock happened before and has been investigated.The most probable root cause for the knob tighten abnormal could be acme nut lead in thread broken for the reason that rotating to lose the knob anti-clockwise rapidly, or rotating the knob leaner or too much strength used, which caused acme screw misaligned with the acme nut lead in thread, acme nut lead in thread broken, then the knob could not be tighten, and link arm could not be tightened.There were no ncmrs, rework, or deviations documented for the reported batch, based on above review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACROBAT-I STABILIZER Z
Common Device Name
STABILIZER, HEART
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key22875057
Report Number2242352-2025-0000962
Device Sequence Number10386645
Product Code MWS
UDI-Device Identifier00607567100008
UDI-Public(01)00607567100008(10)3000464148(17)270318
Combination Product (Y/N)N
Initial Reporter CountrySP
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOM-10000Z
Device Catalogue NumberOM-10000Z
Device Lot Number3000464148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2025
Supplement Date Received by Manufacturer10/14/2025
Initial Report FDA Received Date08/22/2025
Supplement Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN.
Patient Age70 YR
Patient SexMale
Patient Weight65 KG
-
-