DEPUY ORTHOPAEDICS INC US MTO KIN V1 EMSYS OFF CUP IMP; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
299966809 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the item was cross threaded while assembling during testing.The issue was observed in pre-surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.It was reported that there was no further information regarding the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that the item- 200015005 - actual product (299966809 kincise v1 emphasys offset cup impactor mto) was cross threaded while assembling during testing.The issue was observed in pre-surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.It was reported that there was no further information regarding the issue.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device revealed that the adaptor of the mto kin v1 emsys off cup imp was slightly deformed.Additionally, moderate wear consistent with repeated use was observed.No damage on the threads was observed.Although the reported condition was not found, the reported allegation can be confirmed as the observed failure mode prevents the device to work as intended in the same manner.The observed condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the mto kin v1 emsys off cup imp would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
|
| |
|