MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a2, a3, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section e1 - telephone number: (b)(6).The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of this review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information.However, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that during implantation of the preloaded monofocal intraocular lens (iol) the iol was tracking on the side of the tip.The cartridge tip was in contact with the eye but no patient injury was reported.No medical or surgical intervention was required.No concerns were raised regarding patient outcome.No further information as provided.
 
Additional Manufacturer Narrative
Additional information: section h3 - device evaluated by manufacturer? yes device evaluation: photographs provided by the customer evaluated.The photos displayed three handpieces; as it cannot be determine which handpiece applies to this complaint, all three were evaluated.The first photograph displayed the handpiece with the plunger rod advanced, bent, and sticking out the side of the cartridge; the plunger rod tip was also bent.The cartridge was torn and damaged.The second photograph displayed a handpiece with the plunger rod advanced and overriding a lens stuck at the cartridge tip.The last photograph displayed a handpiece with the plunger rod advanced and overriding the lens stuck in the cartridge.Due to photographic quality, no further issues were observed.Since no suspect product has been received, no further evaluation was performed.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE IOL WITH TECNIS SIMPLICITYTM DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key22877833
Report Number3012236936-2025-000207
Device Sequence Number14903790
Product Code HQL
UDI-Device Identifier05050474636248
UDI-Public(01)05050474636248(17)270118
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDCB00
Device Catalogue NumberDCB0000280
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2025
Supplement Date Received by Manufacturer09/04/2025
Initial Report FDA Received Date08/22/2025
Supplement Report FDA Received Date09/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-