MAUDE Adverse Event Report: AVANOS MEDICAL INC. AVANOS, BALLARD CLOSED SUCTION SYSTEM; BALLARD CLOSED SUCTION CATHETER, DOUBLE SWIVEL ELBOW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. AVANOS, BALLARD CLOSED SUCTION SYSTEM; BALLARD CLOSED SUCTION CATHETER, DOUBLE SWIVEL ELBOW Back to Search Results
Model Number 99409
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction: h3 additional information-investigation completion: h6 the actual complaint product was not returned for evaluation; however, the reporter provided photographic/videographic evidence; analysis of the photographic evidence confirmed the complaint as reported; the failure mode is currently under investigation.A review of the device history record is not possible as no lot number was provided.The documented udi is based on the stock/product code provided by the reporter; the udi-pi is not available as no lot number was provided.All information reasonably known as of 22 aug 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.(b)(4).
 
Event or Problem Description
It was reported, the flapper valve dislodged and remained in the circuit; the device was in use for less than two days.The nurse found it inside the tube; there was no reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANOS, BALLARD CLOSED SUCTION SYSTEM
Common Device Name
BALLARD CLOSED SUCTION CATHETER, DOUBLE SWIVEL ELBOW
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, mx 84160
MX   84160
Manufacturer Contact
racheal cooper
2710 north ridge dr nw
ste a
grand rapids, MI 
MDR Report Key22880021
Report Number8030647-2025-00046
Device Sequence Number9313767
Product Code BSY
UDI-Device Identifier00609038994093
UDI-Public(01)00609038994093
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number99409
Device Catalogue Number22703-4J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2025
Initial Report FDA Received Date08/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-