| Model Number |
3300TFX |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Chest Pain (1776); Insufficient Information (4580)
|
| Date of Event |
07/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was learned through implant patient registry that a 29mm 3300tfx aortic valve in aortic position was explanted and replaced with a 29mm 11500a resilia aortic valve after an implant duration of 7 years, 8 months due to unknown reason.Reportedly, patient was in recovery post-operative.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional "customer complaint".The information reported may or may not be related to the edwards device.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated sections: b4, b5, b7, g3, g6, h2, h6 (type of investigation, investigation findings, and investigation conclusions).The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: per (b)(4), stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including coronary artery disease, hyperlipidemia.
|
| |
|
Event or Problem Description
|
|
It was learned through implant patient registry and investigation that a 29mm 3300tfx aortic valve in aortic position was explanted and replaced with a 29mm 11500a resilia aortic valve after an implant duration of 7 years, 8 months due to stenosis.The patient presented with chest pain.Reportedly, patient was in recovery post-operative.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated sections: b4, g3, g6, h2, h6 (investigation findings, and investigation conclusions).A device history record (dhr) review was not performed, as no information regarding a device failure mode was provided.Furthermore, there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error (b)(4).Engineering evaluation summary: there is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, a definitive root cause cannot be conclusively determined.
|
| |
|