| Catalog Number |
490102 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample and/or lot number were provided.Further investigation of the complaint is not possible without a sample and/or lot number.The reported defect was unable to be confirmed.The actual defective device is valuable tool in investigating the cause of this incident.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: describe the event exactly as it occurred: pharmacy technician was preparing for the day priming lines for her bags of fluid for compounding and noticed one of the lines had a foreign substance (black in color - possibly ink or glue) in the priming chamber of the bbraun infusomat space pump line she was going to use.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation, the root cause of the reported incident was unable to be determined.The reported defect was unable to be confirmed.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.The retained units were evaluated and passed the internal testing.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|