Visual inspection: during the investigation, scratches on the outer housing of the m.Blue were determined.Permeability test: a permeability test has shown that all components are permeable.During the permeability test bloody liquid were flushed out.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical position.The results show that the valves are operating within the specified tolerances in their respective relevant position.Adjustment test: the valves were tested and both valves are not adjustable in all pressure settings.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection: after dismantling of the valves, organic deposits were found in all components.Results: according to the investigation results, a non-adjustability at the m.Blue and the progav 2.0 were detected.The visible deposits have led to the functional deviations.Furthermore, we can determine visible deposits inside the control reservoir.The deposits had no effect upon the specifications at the time of the examination.The reservoir operated within all specifications.Deposits caused by substances naturally present in the patient's body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve.The examination of the return has been completed with the drafting of this report.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|