• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. STEALTHSTATION¿ S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. STEALTHSTATION¿ S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735667
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3,h6) no parts have been received by the manufacturer for evaluation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information regarding a navigation system.It was reported that there was an inaccuracy during the procedure.
 
Additional Manufacturer Narrative
Medtronic was made aware of multiple occurrences of inaccuracies that occurred on this system.Medtronic was not made aware of these as they occurred, but after the fact.The following reports are related to this event: 1723170-2025-03044, 1723170-2025-03045, 1723170-2025-03046, and 1723170-2025-03149.Servicing was complete and documented in regulatory report # 1723170-2025-03149.H2 additional information: d10 and additional codes updated.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735740 software, serial lot/sw version: 2.1.0 product id: 9736356 solid state drive, serial lot/sw version: - product id: 9735821 camera, serial lot/sw version: - img code g0200702 applies to the solid state drive and g05001 applies to the camera.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
See 1723170-2025-03149 for additional information.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received stating that the site has faced issue in multiple long segment spine fixation cases.Inaccuracy were faced in cases performed on (b)(6) 2025.All of these were cases where fixation was for more than 2/3 levels.There was approximately 5mm in inferior and superior directions, gross inaccuracy in medial and lateral directions.The clinical workflow for l1-l5/s1 procedures were as follows: the navigation system was position on the caudal end.There was adequate exposure of l1-s1 levels.Short clamp and stealthair reference frame on s1 spinous process.The clamp held the spinous process snuggly.They checked for serrations on the clamp and the appeared to be sharp enough to hold the spinous process.There was a low dose or standard 3d spin was taken with a patient in apnea.Immediately post the 3d spin, the surgeon started with screws.They used the orange and grey navlocks with awl sharp, thoracic probe, 4.5mm legacy cannulated tap followed by screws.They have observed the case with the same workflow.Immediately after the spin they asked the surgeon to check the accuracy at each level.They used all three instruments to check the accuracy.At l1-l2 levels there was an error of 2mm to 3mm.They put the instrument at spinous process, lamina, pedicle entry and the instrument was seen to be in air or at different location other than the anatomy.They have observed the farther they go from the reference frame the error was displayed in the navigation.L3-s1 accuracy showed perfect at the same t ime.They were having received complaints from the surgeon mentioning the same described issue in multiple cases of similar long segment fixations.
 
Additional Manufacturer Narrative
H3, h6: the logs were reviewed for (b)(6) 2025 and confirmed the workflow began with select procedure for a spine procedure, followed by multiple switches between instruments and export tasks, then moved to image acquisition, and finally transitioned to navigation after a significant delay.On (b)(6) 2025, the longest transition occurred between image acquisition navigation, indicating the system remained in image acquisition for nearly an hour before moving forward, which is much longer than typical transitions.On (b)(6) 2025, among the three sessions, the first session had the longest delay before navigation.After starting with select procedure and instruments, the workflow moved to image acquisition but did not transition to navigation for an extended period, suggesting prolonged activity or waiting time during image acquisition before proceeding to navigation.On (b)(6) 2025, the first session began with select procedure and ended without further transitions, likely indicating initialization or reset.The second session moved through instruments and image acquisition, and involved multiple back-and-forth transitions, suggesting repeated adjustments.Navigation occurred after a moderate delay in image acquisition.Later, a back_task returned to image acquisition, followed by a switch to export, and finally back to select procedure, indicating session reset or preparation for a new procedure.Overall, (b)(6) shows moderate delays and high task churn, likely due to user-driven corrections or imaging complexity.There was no archive shared by the user in the log attachment.No log evidence or system errors were found to explain the auto-registration inaccuracy.Based on available data, anatomical movement relative to the reference frame during screw placement is suspected but remains unconfirmed, as such events are not captured in logs or archives.There is insufficient information to determine whether a software anomaly contributed to the issue.Codes b01, c19, and d14 are applicable to this analysis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key22882443
Report Number1723170-2025-03043
Device Sequence Number8861240
Product Code HAW
UDI-Device Identifier00763000359782
UDI-Public(01)00763000359782(11)210309
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K162309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9735667
Device Catalogue Number9735667
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/19/2025
Supplement Date Received by Manufacturer08/19/2025
11/10/2025
01/16/2026
Initial Report FDA Received Date08/25/2025
Supplement Report FDA Received Date09/05/2025
11/11/2025
01/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11...; SEE H11....
Patient SexUnknown
-
-