Additional information: d9, g3, h3, h6.The complaint allegation is confirmed.One unpackaged ar-4030c-08 serial/batch number (b)(6) was received for investigation.Upon visual inspection, the bio screw was fractured at the distal end, close to the tip.Also, a crack was noted close to the break.Functional testing was not performed as the device was received damaged.The most likely cause(s) of the reported failure can be user error, such as improper bone preparation, misaligned insertion, and/or prying the device during insertion.
|