• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. Ø8X 30MM BC IF SCRW, VENTED; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. Ø8X 30MM BC IF SCRW, VENTED; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number Ø8X 30MM BC IF SCRW, VENTED
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during arthroscopic suturing of the meniscus the screw broke during insertion.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.The complaint allegation is confirmed.One unpackaged ar-4030c-08 serial/batch number (b)(6) was received for investigation.Upon visual inspection, the bio screw was fractured at the distal end, close to the tip.Also, a crack was noted close to the break.Functional testing was not performed as the device was received damaged.The most likely cause(s) of the reported failure can be user error, such as improper bone preparation, misaligned insertion, and/or prying the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Ø8X 30MM BC IF SCRW, VENTED
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22882530
Report Number1220246-2025-03605
Device Sequence Number21048112
Product Code MAI
UDI-Device Identifier00888867277656
UDI-Public(01)00888867277656(11)240307(17)280229(10)15217387
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K202535
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberØ8X 30MM BC IF SCRW, VENTED
Device Catalogue NumberAR-4030C-08
Device Lot Number15217387
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2025
Supplement Date Received by Manufacturer10/21/2025
Initial Report FDA Received Date08/25/2025
Supplement Report FDA Received Date10/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-