| Catalog Number |
03157334001 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The electrode lot number was 31244547 with an expiration date of 20-mar-2026.The field service engineer inspected the tubing and found it to be okay.He found crystallization on the edges of the snappak, which was replaced.He performed daily cleaning, a conditioning cycle, calibration, and qc.The customer replaced the electrodes.A reproducibility test was performed, and the results were consistent.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable sodium results from the electrolyte analyzer w/o starterkit 9180.Example 1 initial result was 115 mmol/l, and the repeat results were 136 mmol/l and 137 mmol/l.Example 2 initial result was 114 mmol/l, and the repeat results were 115 mmol/l and 141 mmol/l.The questionable results were not reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was unable to determine a specific root cause.However, the issue appeared to be resolved after the service visit.
|
| |
|
Search Alerts/Recalls
|