COOK INC AMPLATZ ULTRA STIFF PTFE DOUBLE FLEXIBLE WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number G53569
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: zip/postal code = (b)(6).E3: occupation: unknown; health professional = yes.H3: device evaluated by mfg = not returned to manufacturer.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
As reported, during ureteroscopy procedure, the coating of the 'amplatz ultra stiff ptfe double flexible wire guide' flaked off into the patient.Additional information regarding the event and patient outcome has been requested.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: as reported, during ureteroscopy procedure, the coating of the 'amplatz ultra stiff ptfe double flexible wire guide' flaked off into the patient.Reviews of documentation including the complaint history, device history record (dhr), quality control (qc) procedures and instructions for use (ifu), were conducted during the investigation.The complaint device was not returned to cook for investigation.Two photos were provided, one shows the product label confirming the lot.The second photo appears to be through a scope and shows light blue debris in the patient's anatomy.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) found no related non-conformance's that could have contributed to the reported failure mode.A complaint history database search showed no other related complaints associated with the failure mode for the complaint device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in the field.Cook also reviewed product labeling.The ifu packaged with the device did not provide any information related to the reported issue.Based on the available information, no product returned, and the results of the investigation, cook has concluded a definitive cause could not be determined from the available information.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZ ULTRA STIFF PTFE DOUBLE FLEXIBLE WIRE GUIDE
Common Device Name
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key22883021
Report Number1820334-2025-01054
Device Sequence Number18551892
Product Code OCY
UDI-Device Identifier00827002535695
UDI-Public(01)00827002535695(17)270621(10)16121247
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K082536
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG53569
Device Catalogue NumberAUS-038145-DF
Device Lot Number16121247
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2025
Supplement Date Received by Manufacturer11/06/2025
Initial Report FDA Received Date08/25/2025
Supplement Report FDA Received Date11/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-