| Catalog Number |
151312110 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inadequate Osseointegration (2646)
|
| Date of Event |
08/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient has a persistent infection post tka utilizing revision components due to extreme deformity.All components were removed and a static spacer was placed utilizing lps components following a thorough debridement.Resorbable antibiotic beads were also placed.Patient has had multiple i&ds following this procedure and ders were submitted for each.Porous components did not show ingrowth, suggesting that the infection has been present throughout care.No allegation was made against any component.Doi: (b)(6) 2025.Dor: (b)(6) 2025.Affected side: right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: b5 and d10.
|
| |
|
Event or Problem Description
|
|
Additional information received states that not loose, just not ingrown.Not the same thing.
|
| |
|
Search Alerts/Recalls
|