| Model Number |
DVFB1D1 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Electromagnetic Interference (1194); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
04/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported via internal review of historical implantable cardioverter defibrillator (icd) data transmitted from the device that inappropriate high-voltage therapy was delivered for an episode detected in the ventricular fibrillation (vf) zone due to noise on the right ventricular (rv) lead channel.The source of the noise was unclear, but due to the regularity of occurrence, it may have been due to electromagnetic interference (emi).The icd and rv lead remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|