| Model Number |
B1070-040 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure performed was to treat mildly calcified, mildly tortuous left superficial femoral artery that is 90% stenosed.The first 7.0x40mm armada 35 balloon dilatation catheter (bdc) over an unspecified guide wire completely failed to inflate.During inflation of a second 7.0x40mm armada 35 bdc, the balloon ruptured during the first inflation at 10 atmospheres.A third armada 35 bdc was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The additional device referenced in b5 is filed under a separate medwatch report number.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information provided, a definitive cause for the reported balloon rupture could not be determined possible factors that could contribute to balloon material ruptures include, but are not limited to, balloon damage during processing of the balloon material, inflation technique, interactions with other devices or patient anatomy.In this case, a conclusive cause for the reported balloon rupture could not be determined since the device was not returned for analysis.D9 correction: device available for evaluation updated from yes to no.
|
| |
|
Search Alerts/Recalls
|