• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INJECTOR LUER LOCK N35C MULTIPACK; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INJECTOR LUER LOCK N35C MULTIPACK; CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM Back to Search Results
Catalog Number 515005
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
This is a complaint about onivide coring.Occurred during onivade preparation.2 vials were prepared.1 vial was returned to the infusion bag after visual inspection of the inside of the vial confirmed that no coring had occurred.The other vial (the one sent to us) had coring and we judged that it was not possible to collect.The company then checked the bag in which the first vial was returned and confirmed that coring had occurred.As a result, two vials were discarded.The injector and protector were connected.(the lot of injector 515005 is 2410010).
 
Additional Manufacturer Narrative
Follow up mdr for device evaluation/correction: based on investigation results, this complaint is no longer reportable.The involved device did not contribute to the reported failure, coring.Device evaluation it was reported that particles were observed inside the vial and iv bag during preparation.For investigation, one protector assembled onto a vial, one iv bag, one l connector, and a spike set not manufactured by bd were provided to our quality team.Inspection of the returned samples confirmed the presence of a gray particle in both the vial and the infusion bag.Visual assessment determined that the particle in the vial originated from the vial stopper, while the particle found in the iv bag was consistent with material from the iv bag stopper.During evaluation, it was observed that the protector needle penetrated the vial stopper at a raised area, resulting in fragmentation of the stopper.A small stopper piece was also found within the protector needle.Additionally, the iv bag stopper was noted to be significantly torn.No damage or defects were identified within our devices that could have contributed to this observation.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing subprocesses according to procedures.A device history review was performed for protector lot 2501110, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this observation.As a lot number for the lconnector was unavailable for this incident, a device history record review could not be completed.Coring of the rubber stopper may occur depending on stopper quality, design, or if a poor connection is made.Based on the investigation and sample evaluation, the vial stopper fragmentation was attributed to the protector needle piercing a raised portion of the stopper during assembly.Due to the extent of damage observed on the iv bag stopper, it was not possible to definitively determine which spike contributed to the observed particulate matter; therefore, a specific root cause could not be identified.No issues related to the injector were identified.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INJECTOR LUER LOCK N35C MULTIPACK
Common Device Name
CLOSED ANTINEOPLASTIC AND HAZARDOUS DRUG RECONSTITUTION AND TRANSFER SYSTEM
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON, S.A.
cr mequinenza
s/n
fraga, huesca 22520
SP   22520
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22885228
Report Number3003152976-2025-00471
Device Sequence Number15666865
Product Code ONB
UDI-Device Identifier00382905151064
UDI-Public(01)00382905151064
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number515005
Device Lot Number2410010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2025
Supplement Date Received by Manufacturer03/22/2026
Initial Report FDA Received Date08/25/2025
Supplement Report FDA Received Date03/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-