RESPIRONICS, INC. TRILOGY 202 VENTILATOR, CANADA; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number CA1040000
Medical Device Problem Code Circuit Failure (1089)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging during preventive maintenance, an error code related to the oxygen blending module was discovered in the log file.There was no harm or injury reported.The device's oxygen blending module board was replaced to address the issue.Additionally, a generic urgent reminder error code was discovered in the log file.This error code is informational.It is a notification to the user that there is a ventilator service required error active, and no specific replacement parts are required for this error message.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 202 VENTILATOR, CANADA
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22885229
Report Number2518422-2025-109338
Device Sequence Number15683084
Product Code CBK
UDI-Device Identifier00606959022799
UDI-Public(01)00606959022799
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K093905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberCA1040000
Device Catalogue NumberCA1040000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/18/2025
Initial Report FDA Received Date08/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/21/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-