H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
It was reported that a patient with a 19mm 3300tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 8 years, 2 months due to stenosis (mg 75).The patient presented with shortness of breath.The tavr was successfully performed with a 20mm 9755rsl valve.The patient was in stable condition post procedure and discharged on pod #11.Per fcs/ew1: valve intentionally deployed above the surgical valve sewing ring to support vector procedure, patient was on ecmo.
|
|
H11: additional manufacturer narrative: updated: g3, g6, h2, h6: (type of investigation, investigation findings, investigation conclusions).The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency."" there is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia and chemotherapy/radiotherapy.
|