• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751767
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that actuation issue of vitrectomy probe occurred during cataract and vitrectomy combined surgery.The condition of aspirating was unknown.The surgery was completed after replacing the product with another one.There was no patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.2., h.3., h.6.And h.11.Upon visual inspection of the probe the shut pneumatic driveline was detached from engine port.It appeared there was insufficient solvent on the tubing end which likely resulted in a weak bond contributing to the customer's experience.The detachment of the pneumatic tubing renders the device inoperable (unable to use) and is typically visually detectable by the user during setup.The investigation conducted a non-conformance review of the reported lot number.No deviation was identified that would have contributed to this event and all corresponding production release specifications defined in the device master record were met.The most likely root cause of the customer's complaint is consistent with a manufacturing error than can occur during the wetting/assembly process of the pneumatic tubing with solvent and subsequent insertion to the probe engine.In order to mitigate the occurrence of this type of event, during the manufacturing process, in-process checks during assembly and after final assembly are utilized to help screen out this type of defect from occurring.Potential contributing factor could be related to improper handling of the product if the user chooses to disengage the rear ergonomic shell, which can pull-on the pneumatic tubing that are bonded to the engine ports.No further investigative or manufacturing actions are warranted at this time, based on our current tracking, there are no adverse trends for this reported complaint and lot.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trends and take further action, as appropriate.Complaint data for all company products is reviewed monthly to monitor adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trends and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22886267
Report Number1644019-2025-03000
Device Sequence Number20068512
Product Code LRO
UDI-Device Identifier00380657517671
UDI-Public(01)00380657517671(17)260430(11)240525(10)16U8RN
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number8065751767
Device Lot Number16U8RN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/31/2025
Supplement Date Received by Manufacturer01/16/2026
Initial Report FDA Received Date08/25/2025
Supplement Report FDA Received Date02/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONSTELLATION VISION SYSTEM.
Patient SexUnknown
-
-