• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH DRILL BIT AXSOS 3 NON-LOCKING, SHORT, Ø2.5 X 216MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH DRILL BIT AXSOS 3 NON-LOCKING, SHORT, Ø2.5 X 216MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 705025S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tap tip broke off and could not be removed.The procedure was terminated with the tap tip remaining in the patient's bone.In same procedure, the drill bit was broken.All debris were removed from the patient.
 
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that the tap tip broke off and could not be removed.The procedure was terminated with the tap tip remaining in the patient's bone.In same procedure, the drill bit was broken.All debris were removed from the patient.
 
Additional Manufacturer Narrative
Correction to h6(device code).The reported event could be confirmed since the device was returned for evaluation and matches the alleged event.The received device shows signs of prior usage, as we can see heavy, scratch marks throughout the periphery of the device caused by usage over time.The breakage surface indicates breakage in a brittle manner caused by the application of excessive torsional force combined with axial movement.The fracture surface doesn¿t show any signs of plastic deformation and is shiny in nature.The outer diameter of drill was verified and was observed within specification.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance to the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the above investigation, the root cause could be determined to be user-related as breakage is caused due to application of a combination of torsional and axial load.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT AXSOS 3 NON-LOCKING, SHORT, Ø2.5 X 216MM
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key22888978
Report Number0008031020-2025-01425
Device Sequence Number8790313
Product Code HWE
UDI-Device Identifier07613327083439
UDI-Public(01)07613327083439(17)290731(10)5840AA
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number705025S
Device Lot Number5840AA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/29/2025
Supplement Date Received by Manufacturer10/03/2025
Initial Report FDA Received Date08/25/2025
Supplement Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-