• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 4.5PCF
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Codes Respiratory Insufficiency (4462); Unintended Extubation (4564); Voice alteration (4906)
Date of Event 01/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that after a few days, the cuff was found to have deflated, which necessitated an urgent cannula change.The patient experienced respiratory distress during the process.
 
Additional Manufacturer Narrative
Additional information: a2, a4, b3, and g3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: b5, g3 and h6.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that after a few days, the cuff was found to have deflated, which necessitated an urgent cannula change.The patient experienced a phonation problem during the process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22889100
Report Number9681384-2025-00921
Device Sequence Number11996796
Product Code JOH
UDI-Device Identifier20884521779713
UDI-Public(01)20884521779713(17)290625(10)24F0884JZX
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K182861
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4.5PCF
Device Catalogue Number4.5PCF
Device Lot Number24F0884JZX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2025
Supplement Date Received by Manufacturer08/21/2025
09/09/2025
Initial Report FDA Received Date08/25/2025
Supplement Report FDA Received Date08/26/2025
09/11/2025
Date Device Manufactured06/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age24 MO
Patient SexUnknown
Patient Weight9 KG
-
-