• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC HEMOSPRAY ENDOSCOPIC HEMOSTAT; HEMOSTATIC DEVICE FOR ENDOSCOPIC GASTROINTESTINAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC HEMOSPRAY ENDOSCOPIC HEMOSTAT; HEMOSTATIC DEVICE FOR ENDOSCOPIC GASTROINTESTINAL USE Back to Search Results
Model Number G21049
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
During an upper endoscopy procedure, the physician used a cook hemospray endoscopic hemostat.It was reported the nurse activated the handle, flushed the channel, and introduced the device down the scope.When pressing the button no co2 [carbon dioxide] came out.They did not change catheters as no co2 came out, so another of the same device was used to complete the procedure.The nurse observed that the handle of the second device that worked was cold when spraying and the hemospray that did not work the "handle was not cold as it normally is." [unable to spray-subject of report].A section of this device did not remain inside the patient¿s body.The patient was not hospitalized and did not undergo prolonged hospitalization due to this occurrence.The patient did not experience a delay or require any additional procedure due to this occurrence.The complainant did not report any adverse effects on the patient due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.A corrective action (capa) was initiated to further investigate device failure due to being unable to spray powder.This device is within the scope of the capa.Prior to distribution, all hemospray endoscopic hemostats are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action (capa) was initiated to further investigate device failure due to being unable to spray powder.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPRAY ENDOSCOPIC HEMOSTAT
Common Device Name
HEMOSTATIC DEVICE FOR ENDOSCOPIC GASTROINTESTINAL USE
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key22890006
Report Number1037905-2025-00611
Device Sequence Number8918873
Product Code QAU
UDI-Device Identifier00827002210493
UDI-Public(01)00827002210493(17)271025(10)W4882758
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K200972
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG21049
Device Catalogue NumberHEMO-10
Device Lot NumberW4882758
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2025
Initial Report FDA Received Date08/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENDOSCOPE - UNKNOWN MAKE OR MODEL
Patient SexUnknown
-
-