| Model Number |
320550 |
| Medical Device Problem Codes |
Leak/Splash (1354); Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The consumer reported "leakage" when taking her injections.Stated, sometimes the medication will run down her stomach stated, when she attaches her pen needles to the pen, she will see "leakage" around the hub.Stated, she does not prime before injections.Lot: 4303034 catalog: 320550 date of event: unknown.Samples: no cl.
|
| |
|
Additional Manufacturer Narrative
|
|
This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|