| Model Number |
PM3542 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the left ventricular set screw was unable to be screwed into the device header.The device was not implanted, and a new device was implanted to resolve the event.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of could not tighten the lv-lead into the device connector was confirmed.Analysis revealed lv-lead insertion was being blocked by the lv-setscrew that was turned down in the connector port such that leads could not be inserted passed it.Analysis revealed the setscrew had been turned down in this position by someone using the torque wrench participating in the implant procedure.In lab testing the setscrew was backed out from blocking lead passage into the port.An lv-lead could then be fully inserted into the lv connector port, and the setscrew now can be tightened down on it securing the lead in the connector.No device anomalies were found.This was a user-related problem.
|
| |
|