| Model Number |
BF-P190 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject device had an image failure.The issue occurred during maintenance.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updates fields: b5, d9, g3, h2, h3, h6 and h11.The device was returned to olympus for inspection.Based on the results of the investigation the customer allegation was confirmed.No image was due to damage on the objective lens, and the insertion tube was worn due to scratches.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|