Model Number E0502-1
Medical Device Problem Codes Device Alarm System (1012); Connection Problem (2900); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 05/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10 concomitant products: e2505-10fr, e2505-10fr 15cm disp f/s suct coag x25 (lot#:s22he015), vlft10gen, vlft10gen ft series energy platformx1 (serial#:unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during tonsillectomy procedure, the suction coagulation device was connected to universal adapter but failed to produce coagulation therapy and alarming occurred.The device was repositioned but it was still the same.A new package was opened and connected it to the adapter but it also did not work.It was then held by a staff member without the adapter to the monopolar port, which resolved the issue.The patient was readmitted five days post-op for additional surgery to resolved tonsil hemorrhage.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the suction coagulation device was connected to universal adapter but failed to produce coagulation therapy and alarming occurred.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN MFG PLAINFIELD
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key22892493
Report Number3008361498-2025-00001
Device Sequence Number8851878
Product Code GEI
UDI-Device Identifier10884524002170
UDI-Public10884524002170
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
PREAMEND
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE0502-1
Device Catalogue NumberE0502-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2025
Supplement Date Received by Manufacturer10/29/2025
Initial Report FDA Received Date08/26/2025
Supplement Report FDA Received Date10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient Age42 YR
Patient SexMale
-
-