Product complaint # (b)(4) h3 investigation summary: the product was returned to ethicon for evaluation.Twelve unopened overwrap packets were received for analysis.Product code 8206h is double armed.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.H6 component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please clarify, with the second product involved, when was the needle pulled off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.Please provide the source or name and title of external person providing answers to follow-up.(external person submitting answers to sales rep) additional h11: was surgery delayed due to the reported event? no, action taken when event occurred? the suture was swapped out for the same suture of a different lot number, was procedure successfully completed? yes, were fragments generated? no, if yes, were they removed easily without additional intervention? unknown, if no, explain: the suture was discarded and the same suture of a different lot number was used, patient status/ outcome / consequences no, patient consequence description/was there a clinical outcome experienced by the patient (infection, inflammation, etc.)? no patient consequences, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: no, is the patient part of a clinical study unknown, (b)(4), device property of none, device in possession of none.
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected data: g3.Date received by manufacturer: 7/29/2025.
|