|
It was reported that bd saf-t-intima w/y adapter bl 22ga x.75in, separation adapter from tubing.Started iv on patient with a 22 g saf-t-intima.When clamping the tubing after flushing, the entire tubing piece came disconnected from the blue y-site.The iv had to be removed and another iv placed so the patient could receive treatment.
|
|
Dhr review: the complaint lot# is 4354067, sku is 383323, assembly in suzhou plant on 2025.Jan.6, a planned lot quantity is (b)(4) ea.Review the in-process test record and outgoing test report, all test results meet the product specifications, no abnormal found.Review the product assembly record, no non-conformities, deviations, or rework activities for this lot.Returned sample analysis: no picture or sample provided.Retain sample analysis: sampling 2ea from the retain sample of the same lot to check component assembly status and do leakage test, test result is within product specification.Possible cause analysis: based on the information, tubing and adapter are separated during using.The possible reasons for this type of failure may including: 1.Poor swaging performance between tubing and adapter.2.Extension tubing was damaged.Current manufacture already has control procedures as below to monitor and prevent this kind of defect: 1.Both in-process and outgoing sampling will check the pull force related extension tubing and adapter.2.Both in-process and outgoing sampling check do leakage test for component of extension tubing and adapter.Conclusion(s): there is no picture or defect sample provided, and the retain sample test result is pass, so the root cause is not clear for this issue.
|