• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. MCRYL+ UD 18IN 4-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. MCRYL+ UD 18IN 4-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number MCP496G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Wound Dehiscence (1154)
Date of Event 08/20/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.The diagnosis and indication for the index surgical procedure? on what tissue was the suture used? what was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? how was the suture originally tied (multiple knots, square knot, etc.)? did the wound dehis post-op? if yes, what tissue dehisced? is it known how the wound dehisced? please provide further clarification on the ¿sutures opening on all three incision sites¿.Did the suture break post-op? if so, where was the break noted (termination, middle, end)? did the suture pull out of the tissue? did the knots untie post-op? were there any patient stress factors that precipitated this event? what do ¿particles¿ refer to in the statement ¿leaving particles everywhere¿? did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement or during any re-operation? other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? will product be returned? if so, please provide the return date and tracking information.
 
Event or Problem Description
It was reported that a patient underwent a robot assisted laparoscopic hysterectomy on (b)(6) 2025 and suture was used.The procedure went well and the patient tolerated everything appropriately.The next day on (b)(6) 2025, the on call ob got called to the bedside for this patient due to the suture opening on all three incision sites, leaving particles every where.The site was cleaned and closed again with a suture and a topical skin adhesive.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H6 component code: g07002 no device problem.Additional h3 investigation summary: the product was returned to ethicon for evaluation.One open box with six unopened foil packets were received for analysis.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand and no issues related to breakage sutures or anomalies were observed during the evaluation.A functional test was performed using instron equipment and the tensile force was above the minimum requirements.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCRYL+ UD 18IN 4-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22895841
Report Number2210968-2025-09812
Device Sequence Number5986795
Product Code GAM
UDI-Device Identifier10705031046597
UDI-Public(01)10705031046597(17)270430(10)108H5P
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K050845
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMCP496G
Device Lot Number108H5P
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2025
Supplement Date Received by Manufacturer09/12/2025
Initial Report FDA Received Date08/26/2025
Supplement Report FDA Received Date09/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-