| Model Number |
GIF-1TH190 |
| Medical Device Problem Codes |
Poor Quality Image (1408); Residue After Decontamination (2325)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the gastrointestinal videoscope had blurry vision.It is unknown when this malfunction occurred.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h3, h4.The device was returned to olympus for inspection, and the reported failure was not confirmed.In addition, the following reportable malfunction was identified during the device evaluation: white residue in auxiliary water channel at distal end.Based on the results of the investigation, the definitive root cause of the reported issue could not be determined.However, it is likely the use of products that contain silicone led to the malfunction.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|