The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Initial reporter name:(b)(6).Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that there was a balloon rupture after inserting a urinary foley catheter, urine was found to be leaking out from the connection between the catheter and the balloon.As this was not a temporary occurrence, the catheter was removed and a new one was inserted, and the surgery proceeded without further issues.
|
|
The reported event is confirmed manufacturing related.This is addressed by capa 11092653.Visual evaluation of the returned sample noted one opened (without original packaging), used silicone foley catheter with syringe.Visual inspection noted no obvious observations.During testing, when inflating the catheter balloon with 10ml methylene blue solution (3 drops 1percent aq methylene blue per 100ml distilled water), lumen-to-lumen leakage was observed.This is out of specification per inspection procedure ip7603444, revision 0, which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe." the possible root causes for this failure, per capa 11092653, are funnel wall thickness to thin due to molding core pin shape or extruded tube diameter with variation causing leak in catheter funnel molding.Risk review, labeling review, and dhr review are not required as event has already been investigated per capa 11092653.Refer to capa 11092653 for further details.Correction: d,f,h udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that there was a balloon rupture after inserting a urinary foley catheter, urine was found to be leaking out from the connection between the catheter and the balloon.As this was not a temporary occurrence, the catheter was removed and a new one was inserted, and the surgery proceeded without further issues.Per additional information received via ibc on 28ug2025, stated that the overview of the event was changed to urine leakage from balloon breakage.
|