| Model Number |
1012270-12 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a de novo lesion in the distal right coronary artery (drca) with 90% stenosis, moderate calcification and moderate tortuosity.The 2.0x12mm mini trek balloon was being used for pre-dilatation and was prepared without issues.Resistance with the anatomy was noted during advancement and the balloon ruptured during the first inflation at 8 atmospheres (atms).An unspecified trek balloon was used to complete the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the moderately tortuous and moderately calcified anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|