Investigation evaluation: the product said to be involved was returned in an open pouch.Attached to the device was a small barcode label from the lot number provided in the report.Our laboratory evaluation of the product said to be involved confirmed the report.A functional test could not be performed due to the condition of the returned device.A close visual examination identified a rupture in the balloon material.The rupture is located near the middle of the balloon, and no portion of the device or balloon material appear to be missing.An unknown reddish substance was also observed on the balloon.No other anomalies were detected with the device.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the returned device confirmed the report due to a rupture/split in the balloon material.A definitive cause for the reported observation could not be determined because the condition of the product said to be involved prohibited a complete evaluation.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.A leakage/damage to the balloon material can occur if the balloon comes into contact with a sharp object or a burr in the endoscope channel.Another possible contributing factor to a leakage in the balloon material is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.Prior to distribution, all hercules 3 stage balloons esophageal are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment, no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
|
During an esophagogastroduodenoscopy (egd), the physician used a cook hercules 3 stage balloon esophageal.It was reported [that] the balloon popped at about 4 atm when inflated.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|