| Model Number |
LS1037 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: ls1037, ls1037 ligasure atlas handswitch37cm (lot#:s24c0022), ls1037, ls1037 ligasure atlas handswitch37cm (lot#:s24c0022), vlft10gen, vlft10gen ft series energy platformx1 (serial#:unknown) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure, the device was used and pulled out through the trocar.The device has several cables and they pulley ripped.Another device was used but had the same problem.One handpiece was returned back to stock.There was a small pieces of the broken cable fell into the patient¿s cavity.Some pieces were taken out, but some pieces were too small and couldn¿t be retrieved fully.The procedure was extended by more than 30 minutes and no any medical intervention/surgical intervention to retrieve the broken piece.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection found that the jaw wire's insulation was damaged.One of the jaw wires was severed.Functional testing noted that the device's jaw opening and closing mechanism was functioning properly.The weld integrity of the device was inspected and was found to be within specification.The knife cut of the device was tested on a silicone test strip with acceptable results.The heel gap of the device was measured and it was found to be within specification.The device was detected on the lab generators.The severed wire prevented the device from activating when activation attempts were made.It was reported that a component had disengaged, the device was difficult to remove from the trocar, and a wire was exposed.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: use the appropriately sized trocar to allow for easy insertion and extraction of the instrument.Carefully insert and withdraw the i nstrument through the cannula to avoid damage to the device or injury to the patient.Close jaws using device lever before insertion/extraction in the trocar.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|