| Model Number |
TRIEEA31MT |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during an endometriosis procedure, after firing the stapler, the anvil did not bend after firing.The surgeon replaced the device and used another device to complete the procedure.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection of the staple guide noted the presence of a full complement of staples.Further inspection noted that the green bar was not visible in the indicator window and the safety feature was engaged.The staple guide was observed to be attached to the instrument with no damage to the staple guide.It was reported that the anvil did not bend after firing.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: tilted anvil with no cut.Visual inspection noted that the anvil of the instrument was observed to be tilted and attached to the instrument.Further inspection of the anvil cutting ring showed no traces of knife impression and the ring was received intact.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: contraindications do not use the stapler if unusual effort is required to turn the twist knob in order to visualize at least a portion of the green bar in the indicator window.Also, to make certain the space between the cartridge and anvil is closed and ensure that the green bar is visible in the indicator window prior to firing the stapler.The safety will not release if the green bar is not visible in the indicator window; and when firing the stapler, make sure to fully squeeze the instrument handle until the metal underside of the handle contacts the stapler body to the fullest extent.Caution includes that failure to squeeze the instrument handle fully during firing (partial firing) may result in unacceptable staple formation and/or an incomplete knife cut.This may result in leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|