| Model Number |
PCF-H180AI |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: for the failures "aw-cylinder (tube) has foreign objects", "aw-tube has foreign objects" and "j-tube has foreign objects", the most probable cause is cause traced to known inherent risk of device, which indicates that the reported adverse event is known and documented in the labeling and all reasonable mitigation steps have been taken (including both short or long term known complications or adverse reactions).In the instructions for use (ifu) it is stated: ¿nothing other than sterile water should be used for air/water feeding.No additives should be put into the sterile water.Non-sterile water may cause patient cross-contamination and/or infection.¿ should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the colonovideoscope exhibited foreign matter in the auxiliary water channel/tube (j-tube), and the air/water tube.There was no patient involvement.
|
| |
|