MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE HUMERAL STEM PRIMARY PRESS FIT 17MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE HUMERAL STEM PRIMARY PRESS FIT 17MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 300-01-17
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
B3: date of event unknown.D6b: explanted date unknown.D10: 314-02-34 - uhmwpe post aug glenoid large, right: (b)(6), 310-01-53 - equi, humeral head short, 53mm (beta): (b)(6), 300-10-45 - equi replicator plate 4.5mm o/s: (b)(6), 300-20-02 - equi square torque define screw drive kit: (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that the patient underwent an initial total shoulder replacement on the right side.Subsequently, the patient underwent a revision procedure.It was reported that the glenoid was completely eroded and came out in fragments.All pieces of the implant that remained were removed and the area was thoroughly debrided.The surgeon had considered keeping the stem, but upon inspection it was loose and came out with minimal effort.The patient was revised to a competitor's devices.The patient was last known to be in stable condition following the event.Images and x-rays were provided.No devices were not returned for evaluation.No other information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE HUMERAL STEM PRIMARY PRESS FIT 17MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key22898986
Report Number1038671-2025-02842
Device Sequence Number12974790
Product Code KWS
UDI-Device Identifier10885862079343
UDI-Public10885862079343
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K061454
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2023
Device Catalogue Number300-01-17
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2025
Initial Report FDA Received Date08/26/2025
Date Device Manufactured04/29/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-