• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Catalog Number VBHR090702A
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Codes Pain (1994); Ischemia Stroke (4418)
Date of Event 08/15/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of manufacturing records verified that the lot involved in this complaint met all pre-release specifications.The device remains implanted and was, therefore, not available for analysis.No clinical images enabling direct assessment of product performance were returned for evaluation.Cause of the reported event cannot be established based on the information reported to gore.Further information regarding this event was requested by gore, but no further information has been reported, therefore this investigation is considered complete.Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2025, a patient was treated for a thoracic aneurysm using a gore® tag® thoracic branch endoprosthesis system.A gore® viabahn® endoprosthesis (vsx device) was implanted as an extension into the subclavian artery.On an unknown date, the patient reported pain in her lower left leg.While evaluating the pain, an mri showed evidence of multiple small strokes.Reportedly, the patient is recovering.Because the symptoms of the strokes are minimal, no treatment is being given at this time.2203-a used to capture multiple small strokes with the vsx device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
nick lafave
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key22899493
Report Number2017233-2025-06531
Device Sequence Number14949066
Product Code NIP
UDI-Device Identifier00733132645961
UDI-Public(01)00733132645961(17)280424
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P040037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberVBHR090702A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2025
Initial Report FDA Received Date08/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age73 YR
Patient SexFemale
-
-