• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES KII FIOS FIRST ENTRY; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES KII FIOS FIRST ENTRY; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number CFF73
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit has been returned to applied medical for evaluation.A follow-up report will be sent upon completion of investigation.
 
Event or Problem Description
Procedure performed: laparoscopic distal pancreatectomy.Event description: this is a complaint from the hospital; during treatment, the inner seal fragment of the duckbill fell out through the trocar into the abdominal cavity.No patient injury or illness associated with this event.Replaced to a new hospital inventory cff73 and procedure was completed.The investigation report has been requested by the hospital.This event unit will be returned.Intervention: replaced to a hospital inventory new cff73 and the procedure was completed.Patient status: no patient injury or illness associated with this event.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation, along with a black fragment and clear component.Visual inspection confirmed the complainant¿s experience of seal component fragmentation and dislodgment.The black fragment matched the missing portion of the septum, an internal rubber component of the seal.The clear component was identified to be the shield, an internal plastic component of the seal.Based on the condition of the returned unit and description of the event, it is likely the septum fragmentation and shield dislodgement was caused by an asymmetrical instrument that was inserted through the trocar in a non-axial manner.Applied medical¿s instructions for use (ifu) states that, ¿extra care should be used when inserting angular and asymmetrical instruments, such as ¿j¿ hooks and clip appliers.All instruments should be centered axially when inserted through the seal to prevent tearing.¿ applied medical has performed a historical trend analysis and review of production records, and no trend was identified.
 
Event or Problem Description
Procedure performed: laparoscopic distal pancreatectomy.Event description: this is a complaint from the hospital.During treatment, the inner seal fragment of the duckbill fell out through the trocar into the abdominal cavity.No patient injury or illness associated with this event.Replaced to a new hospital inventory cff73 and procedure was completed.The investigation report has been requested by the hospital.This event unit will be returned.Intervention: replaced to a hospital inventory new cff73 and the procedure was completed.Patient status: no patient injury or illness associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KII FIOS FIRST ENTRY
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
ashleigh mendez
22872 avenida empresa
rancho santa margarita, CA 92688
9497135359
MDR Report Key22900381
Report Number2027111-2025-00716
Device Sequence Number12225968
Product Code GCJ
UDI-Device Identifier00607915117658
UDI-Public(01)00607915117658
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K083638
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCFF73
Device Catalogue Number101245501
Device Lot Number1538550
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/26/2025
Supplement Report FDA Received Date12/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-