| Catalog Number |
D128208 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a pentaray nav and it was not flushing during use on the patient.Prior to reinserting the pentaray nav catheter for post-pfa mapping, it was discovered to not be flushing.The catheter was exchanged with a new pentaray nav catheter to continue.There were no errors from the bd alaris pump as the catheter was hand flushed and it would not flush.The correct catheter settings were set on the generator.
|
| |
|
Additional Manufacturer Narrative
|
|
On 25-aug-2025, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's ref.# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a pentaray nav and it was not flushing during use on the patient.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.An irrigation test was performed, and the catheter failed the test due to the irrigation tube being found folded at the tip area.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The irrigation issue reported by the customer was confirmed.The root cause of the irrigation tube folded was traced to the manufacturing process.The instructions for use (ifu) contain the following information: flush the catheter with heparinized saline prior to insertion into the body.Always follow standard practices of using a continuous drip of anticoagulant fluid under pressure through the proximal luer connector when the device is in the body.As part of the johnson & johnson medtech quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#: (b)(4).
|
| |
|
Search Alerts/Recalls
|