| Model Number |
EC34-I10CL |
| Medical Device Problem Code |
Optical Obstruction (3002)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The time of event is unknown.There was no report of patient harm.Video image failure(cloudy ).
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the cmos module with drive pcb cloudy (not clear view).Based on the result, we concluded that it was caused due to the moisture condensation in the cmos module with drive pcb.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|
| |
|
Search Alerts/Recalls
|