MAUDE Adverse Event Report: KATECHO, INC. PADPRO; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KATECHO, INC. PADPRO; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Catalog Number 2001Z-C
Medical Device Problem Code Failure to Deliver Shock/Stimulation (1133)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Event or Problem Description
The customer reported that the device, 2001z-c, adult/child = 10 kg radiotransparent electrode, zoll, was being used during a transcutaneous pacing procedure on (b)(6) 2025 when it was reported, ¿zoll defibrillator failed to transcutaneously pace in 3 instances, (location).Only connection discovered so far is same lot number for defib pads¿.There was no report of injury, medical intervention, or hospitalization of the patient.When asked the reporter gave further information stating, "i would say it was a normal case that turned into an emergency.We have them on ¿just in case¿ and then, when we need them, they are there.So, in our case, we did a normal outpatient and when we needed them as a back up, that is when it failed." this report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Event or Problem Description
The customer reported that the device, 2001z-c, adult/child = 10 kg radiotransparent electrode, zoll, was being used during a transcutaneous pacing procedure on (b)(6) 2025 when it was reported, ¿zoll defibrillator failed to transcutaneously pace in 3 instances, (location).Only connection discovered so far is same lot number for defib pads¿.There was no report of injury, medical intervention, or hospitalization of the patient.When asked the reporter gave further information stating, "i would say it was a normal case that turned into an emergency.We have them on ¿just in case¿ and then, when we need them, they are there.So, in our case, we did a normal outpatient and when we needed them as a back up, that is when it failed." this report is being raised due to the reported malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
Examination of the returned unopened/unused device 2001z-c found that there was some electric continuity issues.Received 369 of the 2001z-c adult/child 10 kg radiotransparent electrode, zoll in unopened original packaging.Sampled 29 devices and functionally inspected to find some defects.Testing the devices with the fluke 114 multimeter found 4 out of the 29 did not have electric continuity through the device.The other 25 did have electric continuity.A root cause cannot be determined; however, based upon evaluation the risk document a possible cause of this event could be that the substrate material is not conductive enough.A device history record (dhr) review was requested from the manufacturer, and no indication of abnormalities was communicated.A two-year lot history review shows a total of 3 devices for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 7 complaints, regarding 9 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised that misuse (including reuse) of multifunction electrodes, use of compromised or altered product, and/or failure to follow product instructions may result in patient burns, inadequate delivery of therapy, and/or loss of ecg trace quality.We will continue to monitor per regulatory requirements to help ensure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PADPRO
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
KATECHO, INC.
4020 gannett avenue
des moines IA 50321
Manufacturer (Section G)
KATECHO, INC.
4020 gannett avenue
des moines IA 50321
Manufacturer Contact
john berga
11311 concept blvd
largo, FL 33773
7273995358
MDR Report Key22902176
Report Number1320894-2025-00212
Device Sequence Number9089388
Product Code MKJ
UDI-Device Identifier10653405042488
UDI-Public(01)10653405042488(17)280522(10)Y05222504
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P200004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2001Z-C
Device Lot NumberY05222504
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2025
Supplement Date Received by Manufacturer10/07/2025
Initial Report FDA Received Date08/27/2025
Supplement Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-