Model Number 4.23.034
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 05/19/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Primary stability not achieved, implant wasn't completely covered with bone, no thread tap used, immediate implantation, mobility: another size of implant was chosen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEVO RC INICELL, PF 4.5, E4.2, L11.0MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH  2540
Manufacturer (Section G)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH   2540
Manufacturer Contact
astrid weber
neckarsulmstrasse 28
grenchen 2540
CH   2540
MDR Report Key22902349
Report Number3003184527-2025-01735
Device Sequence Number7933031
Product Code DZE
UDI-Device Identifier07640182643492
UDI-Public07640182643492
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K093615
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4.23.034
Device Lot Number34803
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/27/2025
Initial Report FDA Received Date08/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age45 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-