|
Investigation summary: the complaint investigation for discrepant results when using the bd sars-cov-2 reagents for bd max¿ system kit (ref.44500301) lots 2242401 and 2291561 was performed by the review of manufacturing records, review of customer¿s data, and verification of complaints history.Review of the manufacturing records of the bd sars-cov-2 reagents for bd max¿ system kit indicated that lots 2242401 and 2291561 were manufactured according to specifications and met performance requirements.Customer complained about false positive results obtained with the bd sars-cov-2 reagents for bd max¿ system kits.Customer provided pdf runs files of one patient sample (# (b)(6)) and multiple environmental monitoring samples (# (b)(6)) from instrument ct1175 and two other run files (#1 (b)(6)) from instrument ct2083.The pcr curves of these samples showed late amplifications (ct values >35) of n1 or n2 targets without anomaly indicative of true low positive results.Moreover, during a visit at customer site of a bd molecular application specialist for assistance to laboratory cleaning and perform environmental runs, an environmental contamination was found.This contamination could come from another pcr-based instrument (biofire) which is close to the bd max instruments.This was suspected because during environmental monitoring at the site, the customer used a new sample buffer tube which was opened and left on the counter overnight, between the bd max and biofire instruments.Upon testing on the bd max instrument, the following day, a late n1 positive result was obtained (sample tested in run (b)(6), instrument ct2083).Based on the data and information provided, environmental target contamination at customer site is the most likely cause to explain the customer¿s positive results, but the investigation does not suggest any reagents issue.There is no indication of a reagent issue based on the analysis of the complaints received for discrepant results on bd sars-cov-2 reagents for bd max¿ system kit lots 2242401 and 2291561.The root cause was not identified.Bd cannot confirm the complaint based on the investigation that was performed.Bd did not initiate a corrective and preventive action (capa) since no new hazard was identified.Bd quality will continue to monitor for trends.This mdr is being submitted as part of a retrospective review of records dating april 2023-2025, under capa 11910483.
|