| Catalog Number |
8270312 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a cavitron 300 series, 240v uk is alleged that no water coming through handpiece.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: engineer's findings: 'to dismantle unit and untrap internal water tubing as it was caught under the chassis, carried out functionality test and it was found after extensive testing that the stack module would not operate the unit correctly, therefore fitted new item and sync footswitch to base unit and then tested unit again, all seems to be ok.
|
| |
|