| Model Number |
G5A-80A |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device displayed an "ecg fault -501" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including daily/weekly/monthly testing, on/off current testing, patient/circuit impedance testing and shock testing without duplicating the report.The main board will be replaced as a precaution.The device will be recertified and returned to the customer.It is important to note that the device will communicate to the user that it is not rescue ready by providing an auditory beep every 30 seconds and displaying a red rescue ready (nvi) indicator.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|